ST. JUDE MEDICAL FAST-CATH¿ HEMOSTASIS INTRODUCER, CATH-LOCK¿ LOCKING HUB, 12 CM SHEATH, 6 F; INTRODUCER, CATHETER Back to Search Results
Model Number 406701
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 07/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
During a pulmonary vein isolation procedure for atrial fibrillation, the sheath broke and detached during the z-suture and remained in the groin.The patient was transferred to vascular surgery where the detached portion was removed under fluoroscopy.Per the surgeon, there were no further consequences to the patient.
 
Additional Manufacturer Narrative
One 6f fast-cath introducer sheath was received for evaluation.The sheath tubing had been torn and stretched into two sections; both sections were returned; the distal section had been kinked and twisted proximal to the distal tip.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the sheath tubing damage remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH¿ HEMOSTASIS INTRODUCER, CATH-LOCK¿ LOCKING HUB, 12 CM SHEATH, 6 F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15074592
Report Number3005334138-2022-00425
Device Sequence Number12090747
Product Code DYB
UDI-Device Identifier05414734216944
UDI-Public05414734216944
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K061015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number406701
Device Catalogue Number406701
Device Lot Number8446368
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2022
Supplement Date Received by Manufacturer08/30/2022
Initial Report FDA Received Date07/21/2022
Supplement Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-