• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in a shunt in the moderately tortuous and moderately calcified intra-forearm.A 4.0 x 40, 40cm mustang balloon catheter was advanced for dilatation.Inflation was performed at 10 atmospheres for 60 seconds and subsequently at 18 atmospheres for 10 seconds.However, during the second inflation, the balloon ruptured and a pinhole was noted around the center.The device was removed and the procedure was completed with another of same device.No patient complications were reported and the patient was in good condition post procedure.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: a mustang device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from pinhole in the shaft located approximately 8mm proximal from the proximal markerband.There were no leaks from balloon material when inflated and visually no issues.A visual examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A leak test identified a prox bond leak.A visual and tactile examination identified no damage to the tip which may have potentially contributed to the complaint incident.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in a shunt in the moderately tortuous and moderately calcified intra-forearm.A 4.0 x 40, 40cm mustang balloon catheter was advanced for dilatation.Inflation was performed at 10 atmospheres for 60 seconds and subsequently at 18 atmospheres for 10 seconds.However, during the second inflation, the balloon ruptured and a pinhole was noted around the center.The device was removed and the procedure was completed with another of same device.No patient complications were reported and the patient was in good condition post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15075464
Report Number2124215-2022-26536
Device Sequence Number14781011
Product Code FGE
UDI-Device Identifier08714729794103
UDI-Public08714729794103
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103751
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/27/2025
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0029301944
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/15/2022
Supplement Date Received by Manufacturer08/19/2022
Initial Report FDA Received Date07/21/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE USED: RADIFOCUS; GUIDEWIRE USED: RADIFOCUS; INFLATION DEVICE USED: ENCORE 26; INFLATION DEVICE USED: ENCORE 26; INTRODUCER SHEATH USED: 5F; INTRODUCER SHEATH USED: 5F
Patient SexUnknown
-
-