As reported, during a ureteroscopy involving ureteral stent placement, the tip on a amplatz stiff ptfe fixed core wire guide separated and remained inside the patient's kidney.The wire guide was inserted through the ureter along with an unknown manufacturer's semirigid ureteroscope and advanced to the kidney.An unknown manufacturer's ballistic lithoclast was used to break up the stone.Upon removal of the guide wire under x-ray visualization, a 2 to 3 centimeter section of the wire guide separated and remained in the renal pelvis.Attempted removal of the separated section was unsuccessful, and there is no information on a follow up procedure at this time.Additional patient/event information has been requested.
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Event summary as reported, during a ureteroscopy involving ureteral stent placement, the tip on a amplatz stiff ptfe fixed core wire guide separated and remained inside the patient's kidney.The wire guide was inserted through the ureter along with an unknown manufacturer's semirigid ureteroscope and advanced to the kidney.An unknown manufacturer's ballistic lithoclast was used to break up the stone.Upon removal of the guide wire under x-ray visualization, a 2 to 3 centimeter section of the wire guide separated and remained in the renal pelvis.Attempted removal of the separated section was unsuccessful, and there is no information on a follow up procedure at this time.Additional patient/event information has been requested.Investigation - evaluation reviews of the complaint history, device history record, instructions for use, and quality control procedures and a visual inspection of the device were conducted during the investigation.The device was returned in labelled open packaging for investigation.The wire appeared to have been pulled and broken off at the distal end at approximately 145.3 cm from proximal end.There also appeared to be wear marks where the coating has been rubbed off at both the proximal and distal ends.A document-based investigation evaluation was performed.No related non-conformances were recorded.The device history record review provides objective evidence that the device was manufactured to specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was functionally inspected by quality control and no notable gaps in production or processing controls were noted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.The device is provided with instructions for use which caution, "manipulation of wire guide requires appropriate imaging control.Use caution not to force or over manipulate the wire guide when gaining access.¿ the instructions also caution, ¿when using wire guide through a metal cannula/needle, use caution as damage may occur to outer coating.¿ based on the available information, cook has concluded that a cause the complaint could not be established.It was not possible to rule out inadvertent user/procedural issues; manipulation may have been required, as it was reported that the patient had a ¿large¿ stone.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per a review of risk documentation, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|