• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP THORACIC PEDICLE FEELER; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP THORACIC PEDICLE FEELER; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 6002-350-000
Medical Device Problem Codes Fracture (1260); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
Per the customer, the tip of a thoracic pedicle feeler broke off inside the patient and was unable to be removed.The device fractured at the patient's sacroiliac joint during a revision surgery of an iliac bolt that had been previously placed in the patient.Per the clinical specialist observing the case, the thoracic pedicle feeler was positioned directly next to the iliac bolt.While attempting to create a pilot hole for the thoracic pedicle feeler, the iliac bolt was struck, the gear shaft bent and the tip broke off as well as the handle.The procedure was completed successfully without a clinically significant delay; no adverse consequences or medical intervention were reported.
 
Additional Manufacturer Narrative
Device evaluation: follow-up report submitted to document the device evaluation.
 
Event or Problem Description
Per the customer, the tip of a thoracic pedicle feeler broke off inside the patient and was unable to be removed.The device fractured at the patient's sacroiliac joint during a revision surgery of an iliac bolt that had been previously placed in the patient.Per the clinical specialist observing the case, the thoracic pedicle feeler was positioned directly next to the iliac bolt.While attempting to create a pilot hole for the thoracic pedicle feeler, the iliac bolt was struck, the gear shaft bent and the tip broke off as well as the handle.The procedure was completed successfully without a clinically significant delay; no adverse consequences or medical intervention were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THORACIC PEDICLE FEELER
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM   D-79111
Manufacturer Contact
colette chung
1941 stryker way
portage, MI 49002
2693237700
MDR Report Key15076327
Report Number3015967359-2022-01329
Device Sequence Number5338887
Product Code HAW
UDI-Device Identifier04546540503824
UDI-Public04546540503824
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K012380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number6002-350-000
Device Catalogue Number6002-350-000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/24/2022
Supplement Date Received by Manufacturer09/19/2022
Initial Report FDA Received Date07/21/2022
Supplement Report FDA Received Date09/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-