ENCORE MEDICAL L.P. ENCORE SHOULDER; ENCORE REVERSE SHOULDER HUMERAL STEM, STD, SZ8X108MM Back to Search Results
Model Number 530-08-108
Health Effect - Clinical Code Pain (1994)
Date of Event 07/13/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery due to instability and pain.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as instability and pain.The previous surgery and the surgery detailed in this event occurred 3 months and 2 weeks apart.The actual in vivo time for the main part of the complaint is 3 years and 11 months from the primary surgery.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was a ncmr# 44214 associated with the main part # 530-08-108, altivate reverse, humeral stem, standard shell, sz 8x108mm, which documents that out of 15 parts lot, 9 parts were rejected due to porous coating damaged around suture holes.Later the rejected parts were accepted to use as is after proper justification.All other items in the lot were met with the design, fit and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to instability and pain.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Additional Manufacturer Narrative
Corrected data: see d.6a.Manufacturer narrative: additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE SHOULDER
Common Device Name
ENCORE REVERSE SHOULDER HUMERAL STEM, STD, SZ8X108MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key15077238
Report Number1644408-2022-00948
Device Sequence Number12122398
Product Code PHX
UDI-Device Identifier00888912168304
UDI-Public(01)00888912168304
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K141990
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/06/2024
Device Model Number530-08-108
Device Catalogue Number530-08-108
Device Lot Number406T1256
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2022
Supplement Date Received by Manufacturer07/21/2022
Initial Report FDA Received Date07/21/2022
Supplement Report FDA Received Date07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/10/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
506-03-118 LOT 832C1457; 506-03-118 LOT 832C1458; 506-03-126 LOT 834C1468; 508-36-103 LOT 871C1186; 509-01-036 LOT 382P1046
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-