• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ENDOTRACHEAL TUBE WITH HOLDER AND CONNECTOR; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ENDOTRACHEAL TUBE WITH HOLDER AND CONNECTOR; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 100/179/055
Medical Device Problem Codes Inflation Problem (1310); Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the pre-use check, leakage of air from the cuff was detected.No patient injury or adverse affects.No additional information is available for this complaint.
 
Additional Manufacturer Narrative
H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Photos received: six photos were included in agile attachments for evaluation; photos showed a damage on tip of the tube.Samples received: two samples were received in used conditions without original package.Visual inspection results: no defects were found.Leak testing: the returned samples were connected to a machine to identify any leak.Results: gas leakage was detected, the complaint was confirmed.Customer complaint notification was performed on 25-jul-2022 to personnel to make them aware of the reported failure mode.The cause of the reported problem was traced to the manufacturing process.A dhr (device history review) was performed and no problems or issues were identified during this dhr review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ENDOTRACHEAL TUBE WITH HOLDER AND CONNECTOR
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
7-1-1 akasaka,
minneapolis, MN 55442
MDR Report Key15077506
Report Number3012307300-2022-13895
Device Sequence Number14780944
Product Code BTR
UDI-Device Identifier15019315022098
UDI-Public15019315022098
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K790312
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2026
Device Model Number100/179/055
Device Catalogue Number100/179/055
Device Lot Number4169638
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2022
Supplement Date Received by Manufacturer10/21/2022
Initial Report FDA Received Date07/21/2022
Supplement Report FDA Received Date11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/30/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-