Model Number MD170126024S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Extravasation (1842)
Date of Event 06/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The involved device was not returned for evaluation.One photograph was provided showing a split in the lumen approximately 5 cm in length.The exact location of the split cannot be determined from the photograph.Requests for additional information, including the length of time the device was implanted, were not successful.The contract manufacturer conducted a review of the manufacture records for the lot number provided.Their investigation revealed the device was manufactured and inspected according to specification with no non-conformances or abnormalities.The manufacture process includes a 100% leak test performed as a last step in the process.This test would have detected any leaks, holes, or weak spots if it had existed at the time of manufacture.Without an evaluation of the device a root cause cannot be determined.Flushing against resistance is a common cause for lumen breakage.The instructions for use (ifu) contains the following warnings and precautions: do not use infusion equipment which can exceed the working pressure of 1.0bar max/750mmhg (14.5 psi).Only use infusion equipment complying with standards, which do not exceed a shut-off pressure of 1.0bar.Bolus injections should be slow and must not exceed the maximum bolus pressure small syringes will generate excessive pressure and may damage the catheter.The use of 10cc or larger syringes are recommended.Warning: do not flush against resistance.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
An end user reported an issue with a vascu-picc tls pediatric 2.6fx50cm dl mst kit, that was placed into a patient in the neonatal intensive care unit (nicu).Three days after picc placement, the nicu noticed a right femoral extravasation of 61%.When ir arrived to remove the line and place a new one, a longitudinal split along the catheter was noted.The device related failure was reported to have led to patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.6F X 50CM DUAL VASCU-PICC W/MST
Common Device Name
VASCU-PICC
Manufacturer (Section D)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
Manufacturer (Section G)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
Manufacturer Contact
lynn winkler
1499 delp drive
harleysville, PA 19438
2152564201
MDR Report Key15078887
Report Number2518902-2022-00044
Device Sequence Number11756793
Product Code LJS
UDI-Device Identifier00884908147230
UDI-Public884908147230
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K120981
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model NumberMD170126024S
Device Catalogue NumberMD170126024S
Device Lot NumberMQDB120
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2022
Initial Report FDA Received Date07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age8 WK
Patient SexMale
-
-