The reported event of a yellow wrench advisory (controller fault) alarm was not confirmed.A review of the log file spanned approximately 11 days (b)(6) 2022 ¿ (b)(6) 2022 per the timestamps).There were no fault alarms related to the system controller operation presented in the log file.The driveline was disconnected on (b)(6) 2022 at 03:42 - 03:43 for a controller exchange.No other notable alarms were active in the log file.The returned system controller, serial (b)(4), was functionally tested and found to function as intended.The provided information indicated that the patient noted a yellow wrench alarm on the controller that resolved after the controller was exchanged.However, when the patient entered the clinic, the controller was plugged in for troubleshooting, but no fault alarms were noted.No pictures or additional documents related to the reported event were provided to this complaint.As a result, this analysis cannot conclusively determine a root cause for the reported event.The device history records were reviewed and the records revealed the controller (b)(4) was manufactured in accordance with mfg and qa specifications.The system controller was shipped to the customer on 27 dec 2021.Heartmate 3 instructions for use section 7 ¿ ¿alarms and troubleshooting¿ and heartmate 3 patient handbook section 5 ¿ ¿alarms and troubleshooting¿ explains how to properly interpret and troubleshoot all alarms.Heartmate 3 instructions for use section 8 ¿ ¿equipment storage and care¿ and heartmate 3 patient handbook section 5 section 6 ¿ ¿caring for equipment¿ explain how to properly care for the equipment.The patient handbook cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.The manufacturer is closing the file on this event.
|