Model Number M400 HD
Device Problems Break (1069); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/30/2022
Event Type  malfunction  
Manufacturer Narrative
Permobil ab were provided photo evidence of the device when the event occurred.Photos confirmed the seating system having detached from the chassis.Photos appear to indicate a material break having occurred where the spline portion of the elevator is attached to the mounting plate which secures the seating.Permobil ab has requested the suspect component be returned to permobil ab for further analysis in effort to determine potential root cause for the failure.Review of complaint data indicates this to be the first reported failure of this type.Upon receipt of any new information, a follow-up report will be submitted.The dhr was reviewed, and the device was found to have met specification prior to distribution.
 
Event Description
Permobil ab received report claiming as the end-user was driving along a sidewalk, the seating was reported to have detached from the chassis which resulted in the end-user falling to the ground.No injuries were reported to have occurred as a result.
 
Manufacturer Narrative
Permobil ab sent the failed component to an external company that carried out a material and failure analysis.Report results indicated the cause of the component failure was material fatigue which originated from failure to fully bond during the welding process.The fatigue failure was started at several points at bond failure, where the cracks have grown the same and continued until the remaining surface could not cope with the load, which finally resulted in a residual failure.Review of complaint data indicates this to be the first reported failure of this type since the implementation of this component into the market in 2013.Permobil will continue to monitor any similar failures through trend analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERMOBIL M400 HD
Type of Device
POWER WHEELCHAIR
Manufacturer (Section D)
PERMOBIL INC
300 duke drive
lebanon TN 37090
Manufacturer (Section G)
PERMOBIL INC.
300 duke drive
lebanon TN 37090
Manufacturer Contact
kevin bullock
300 duke dr.
lebanon, TN 37090
8007360925
MDR Report Key15079219
MDR Text Key304339341
Report Number1221084-2022-00013
Device Sequence Number1
Product Code ITI
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K123290
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberM400 HD
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/23/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-