| Model Number |
AFFINITI 70 DIAGNOSTIC ULTRASOUND SYSTEM |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The initial investigation found there to be no system malfunction associated with this event.However, an investigation is underway to determine if this manual entry issue can be avoided in the future.Results of the investigation will be included in a follow up report upon its completion.
|
| |
|
Event or Problem Description
|
|
A customer reported their affiniti 70 ultrasound system had produced incorrect volume calculations during a prostate exam due to an incorrect volume calculation entry when composing a custom preset manually.The issue was obvious to the user and there was no harm to the user or patient associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A thorough investigation was performed, including analysis of logs, to identify the root cause of the reported issue.The engineering team confirmed that the system is performing as designed.
|
| |
|