• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® URETHRAL CATHETERIZATION TRAY WITH RED RUBBER CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® URETHRAL CATHETERIZATION TRAY WITH RED RUBBER CATHETER Back to Search Results
Model Number 772415
Medical Device Problem Codes Material Fragmentation (1261); Component Misassembled (4004)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that in and out cath being done on labor patient with epidural to empty bladder.Nursing staff stated that the intermittent catheter was intact when inserted into urethra but came out missing the top portion.It was stated that the patient ended up having to under cystostomy after delivery.As per investigator notification on 06jul2022, it was stated that an additional red rubber catheter was returned for sample evaluation.
 
Additional Manufacturer Narrative
The reported event is confirmed, manufacturing related.One red rubber catheter was returned without the original packaging.The length of the shaft was visually inspected and no lumps, discoloration, or anomalies were noted.Some flashing appeared to be present within the drainage eye.A potential root cause for this event could be, "operator error, dull punch dies" or "machine error".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review is not required as a review of the label could not have prevented the reported event.Corrections: d,f,h h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was evaluated.
 
Event or Problem Description
It was reported that in and out cath being done on labor patient with epidural to empty bladder.Nursing staff stated that the intermittent catheter was intact when inserted into urethra but came out missing the top portion.It was stated that the patient ended up having to under cystostomy after delivery.As per investigator notification on (b)(6) 2022, it was stated that an additional red rubber catheter was returned for sample evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® URETHRAL CATHETERIZATION TRAY WITH RED RUBBER CATHETER
Common Device Name
TRAY WITH RED RUBBER CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15079770
Report Number1018233-2022-05676
Device Sequence Number17346904
Product Code KOD
UDI-Device Identifier00801741030413
UDI-Public(01)00801741030413
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model Number772415
Device Catalogue Number772415
Device Lot NumberNGGN4134
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/06/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date07/21/2022
Supplement Report FDA Received Date12/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-