Model Number A4E01
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We have received the complaint device for evaluation and we have confirmed the reported incident.We observed the balloon of this catheter has ruptured.We were able to flush the inflation lumen and the irrigation lumen properly.We did not observe any other defects in this catheter.Our ifu clearly/appropriately warns the users about the risks that could occur with the use of the embolectomy catheter including the risk of balloon rupture due to exposure to calcified plaque and overinflated balloon.The arterial embolectomy catheter is not recommended for the removal of fibrous, adherent, or calcified material ( eg.Chronic clot, atherosclerotic plaque ).This catheter is not designed to withstand the additional pull force needed to remove these materials.There was no harm to the user because of the balloon malfunction.The surgeon used another catheter to complete the procedure.
 
Event or Problem Description
The balloon of the syntel over-the-wire embolectomy catheter ruptured during an embolectomy procedure.The balloon ruptured during the first pass when it was inflated in the patient's blood vessel.So, the surgeon used another catheter for the procedure.There was no harm to the patient because of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNTEL SILICONE OVER-THE-WIRE EMBOLECTOMY CATHETER
Common Device Name
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
lemaitre vascular, inc.
63 second ave MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer Contact
pragya thikey
63 second ave
burlington, MA 01803
7812212266
MDR Report Key15079782
Report Number1220948-2022-00065
Device Sequence Number14696033
Product Code DXE
UDI-Device Identifier00840663109869
UDI-Public00840663109869
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K022145
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA4E01
Device Catalogue NumberA4E01
Device Lot NumberPYT1066
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2022
Initial Report FDA Received Date07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-