MAUDE Adverse Event Report: SYNTHES GMBH 4.5MM VA-LCP CURVED CONDYLAR PLATE/12HOLE/266MM/LT-STER; CONDYLAR PLATE FIXATION IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 4.5MM VA-LCP CURVED CONDYLAR PLATE/12HOLE/266MM/LT-STER; CONDYLAR PLATE FIXATION IMPLANT Back to Search Results
Model Number 02.124.413S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Type of Reportable Event Serious Injury
Event or Problem Description
Device report from synthes reports an event in great britain as follows: patient sustained periprosthetic fracture (ppf) below femoral nail in march and was fixed with va condylar plate on an unknown date in march 2022.Patient was readmitted approximately 3 months after implant with plate fracture.No new trauma.Had been mobilizing in residential home.Normal to low bmi.Fracture non-united, but no evidence of infection.Revision occurred (b)(6) 2022.This is report 1 of 1 for (b)(4) this report is for a condylar plate this complaint is linked to (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: unknown between march 2022 and june 2022 implant date unknown date in march 2022 complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter is j&j company representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: investigation summary = the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from the photo.Visual analysis of the photo revealed that unk - plates: trauma had system construction is visible at the femur left with signs of bone fracture and the plate was observed broken.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for unk - plates: trauma.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review = product code: 02.124.413s, lot number: 374p884: a manufacturing record evaluation was performed for the sterile lot number, and no non-conformances were identified.H10: correction narrative: d6a: the implantation date was inadvertently missed on the initial report; has been updated to reflect the correct information.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: g1 h3, h6: product code: 02.124.413s lot: 374p884 manufacturing site: mezzovico release to warehouse date: 05 october 2021 expiration date: 01 september 2031 a manufacturing record evaluation was performed for the sterile lot number, and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.¿ visual analysis of the returned sample revealed that va-lcp condylar plate 4.5/5.0 le 12ho l2 has broken in two parts, the broken fragments were returned.No other issues were observed.A dimensional inspection was not performed for the va-lcp condylar plate 4.5/5.0 le 12ho l2 due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.¿ ¿the overall complaint was confirmed as the observed condition of the va-lcp condylar plate 4.5/5.0 le 12ho l2 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.¿ ¿based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.¿ drawing/specifications reviewed? current and manufactured revisions were reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9 h4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM VA-LCP CURVED CONDYLAR PLATE/12HOLE/266MM/LT-STER
Common Device Name
CONDYLAR PLATE FIXATION IMPLANT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK MEZZOVICO (CH)
via cavazz 5
mezzovico 6805
SZ   6805
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15079869
Report Number8030965-2022-05074
Device Sequence Number5101430
Product Code JDP
UDI-Device Identifier10886982042880
UDI-Public(01)10886982042880
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K110354
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number02.124.413S
Device Catalogue Number02.124.413S
Device Lot Number374P884
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2022
Supplement Date Received by Manufacturer08/08/2022
09/06/2022
Initial Report FDA Received Date07/21/2022
Supplement Report FDA Received Date08/30/2022
09/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONNECTION SCREW F/LOCKING ATTACHMENT PL; POSITION-PIN 4.5 F/LCP SST
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexFemale
-
-