ONKOS SURGICAL ELEOS; POROUS, HA TREATED, ROUND, 34MM, 32MM MODULAR POROUS BIOGRIP HA COLLAR
|
Back to Search Results |
|
| Model Number |
PB-3200R-03M |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Insufficient Information (4580)
|
| Date of Event |
06/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient experienced an infection and underwent a revision surgery.The following implants were removed and replaced during the revision: eleos tibial hinge, eleos tibial poly spacer, eleos distal femur axial pin, eleos distal femur and 2 x eleos male-female midsections.No additional information regarding this adverse event has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is in progress.When it is complete a supplemental mdr will be submitted accordingly.Multiple mdrs were submitted for this adverse event: 3013450937-2022-00202, 3013450937-2022-00203, 3013450937-2022-00204, 3013450937-2022-00205, 3013450937-2022-00206, 3013450937-2022-00207, 3013450937-2022-00208, 3013450937-2022-00210, 3013450937-2022-00211, 3013450937-2022-00212 and 3013450937-2022-00213.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient experienced an infection and underwent a revision surgery.The following implants were removed and replaced during the revision: eleos tibial hinge component, eleos tibial poly spacer, eleos distal femur axial pin, eleos distal femur, and two eleos male-female midsections.The following eleos implants remain implanted from the patient's original surgery: cemented segmental stem, biogrip modular collar, collar locking ring, tibial baseplate, canal-filling stem extension, and tibial baseplate screw.No additional information regarding this adverse event has been provided.
|
| |
|
Search Alerts/Recalls
|
|
|