Model Number H7493941828250
Medical Device Problem Codes Device-Device Incompatibility (2919); Device Dislodged or Dislocated (2923); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that stent dislodgement and stent damage occurred resulting in additional intervention.The 70% stenosed target lesion was located in the non-tortuous and mildly calcified right coronary artery (rca).A 2.50 x 28mm synergy xd drug-eluting stent was advanced through a previously implanted 2.50x28 synergy stent for treatment.However, it was noted that the stent was caught on the implanted stent and dislodged from the balloon.The stent became lodged into the ostial rca and both stents were noted to be damaged.The stent was snared and removed from the patient successfully.The procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
(b)(6).
 
Additional Manufacturer Narrative
E1: initial reporter address 1: (b)(6).Device evaluated by mfr.: synergy xd mr us 2.50 x 28mm stent delivery system was returned for analysis.The device was returned without the stent.A review of the manufacturing stent profile data was performed and the stent outer diameter at the time of manufacture was within max crimped stent profile measurement.The balloon cones were reviewed, and the folds were noted to be relaxed and crimp stent markings were visible on the balloon body.The balloon was noted to contain traces of contrast media inside the balloon.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks along the length of the hypotube shaft.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues.No other issues noted with device.
 
Event or Problem Description
It was reported that stent dislodgement and stent damage occurred resulting in additional intervention.The 70% stenosed target lesion was located in the non-tortuous and mildly calcified right coronary artery (rca).A 2.50 x 28mm synergy xd drug-eluting stent was advanced through a previously implanted 2.50x28 synergy stent for treatment.However, it was noted that the stent was caught on the implanted stent and dislodged from the balloon.The stent became lodged into the ostial rca and both stents were noted to be damaged.The stent was snared and removed from the patient successfully.The procedure was completed with another of the same device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15080366
Report Number2134265-2022-07723
Device Sequence Number14466967
Product Code NIQ
UDI-Device Identifier08714729980834
UDI-Public08714729980834
Combination Product (Y/N)Y
Initial Reporter StateOK
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/14/2023
Device Model NumberH7493941828250
Device Lot Number0028250778
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/29/2022
Supplement Date Received by Manufacturer08/17/2022
Initial Report FDA Received Date07/21/2022
Supplement Report FDA Received Date09/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-