| Model Number |
ORT300 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported that during a morning qa, drops of hydraulic fluid were observed at the base of the ort300 operating room table.No issues were reported with respect to hydraulics controlled table functions.There was no patient involvement when the apparent leak was observed.
|
| |
|
Additional Manufacturer Narrative
|
|
An imris customer service engineer addressed the issue on-site.Upon inspection, the cse confirmed presence of a hydraulic fluid leak and traced it to the connection between a ptfe hydraulic line and the trendelenburg cylinder.The fiting securing the connection was adjusted and tightened.The cse measured the hydraulic fluid level in the operating room table and confirmed it was within specification.No components required replacement and no additional fluid was added to the table.The table was tested to confirm function after the issue was addressed; no table functional issues were observed by the cse or reported by the customer.
|
| |
|