| Model Number |
AAB00 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was a sticky consistency on the anterior portion of the intraocular lens.The lens was fully inserted into patient's right eye, however, was removed from the eye and a new lens of same model and diopter was inserted.There was no patient injury.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.The device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: jul 29, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the complaint lens was received cut into three pieces.The lens was inspected with both a dark and a white background.No alleged foreign material could be identified and no "sticky material" was found on the lens.Conclusion: the complaint issue was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|