| Model Number |
23375 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 15-jul-2022.It was reported that balloon damage occurred.The patient presented with peripheral arterial disease.The target lesion was located in the non-tortuous and non-calcified superficial femoral artery.A 12-4/5.8/75 xxl vascular balloon catheter was advanced for dilatation.However, during the procedure, a kink was noted on the balloon itself.The device was removed and the procedure was completed with a non-boston scientific balloon.No patient complications were reported.However, device analysis revealed a balloon pinhole.
|
| |
|
Additional Manufacturer Narrative
|
|
A xxl device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was present in the balloon which is indicative of a leak.The balloon was attached to an encore inflation device, subjected to positive pressure and liquid was observed to be leaking from a balloon pinhole located 2mm proximal from the distal markerband.The rated burst pressure for this device is 8 atmospheres.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual examination identified no issues with the tip of the device that could have contributed to the complaint incident.This concludes the product analysis.
|
| |
|