| Model Number |
CNW0T5 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.The product history record could not be reviewed because facility did not provide a lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that after implantation of an intraocular lens (iol) the edge of iol was found to be deformed after implanting.Although he/she thought that it was damaged by the implanting process, it turned out that it was already deformed when taking it out of the case when checked through surgical video after the surgery.The surgery was completed without product replacement.The product was not replaced since it was a slight deformation and it was considered that there would be no damage to the patient's health.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The video shows lens to be damaged as the lens is removed from the iol case.The root cause is potentially manufacturing related.If the iol becomes misaligned in the lens case, iol damage may occur.Based on the results from the product history record, the products met release criteria.Three photos provided.The optic edge appears to be chipped in the two photos taken during surgery, the reported complaint cannot be confirmed from the remaining the manufacturer internal reference number is: (b)(4).
|
| |
|