| Model Number |
02.12.0004R |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Joint Laxity (4526)
|
| Date of Event |
07/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 08-jul-2022.Lot 160122: 55 items manufactured and released on 05-apr-2016.Expiration date: 2021-02-27.No anomalies found related to the problem.To date, all items of the same lot have been sold with no similar reported case during the period of review.Additional components involved: gmk-sphere 02.07.1204r tibial tray fixed cemented size 4 r (k090988) lot.160458: 34 items manufactured and released on 24-mar-2016.Expiration date: 2021-03-08.No anomalies found related to the problem.To date, all items of the same lot have been sold with no similar reported case during the period of review.
|
| |
|
Event or Problem Description
|
|
At about 6 years from the primary surgery, it was discovered the loosening of femoral and tibial implants.Femur, liner and tibia were removed and a revision prosthesis was put in place.The surgery was completed successfully.
|
| |
|