| Model Number |
B1PC (+20.50 D) |
| Medical Device Problem Code |
Sharp Edges (4013)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Capsular Bag Tear (2639)
|
| Date of Event |
06/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This initial emdr is being submitted to fda as a reportable malfunction that occurred in the usa.Haptic damage is indicated as a potential malfunction related to the iol, as covered under the warnings section of the product's instructions for use (ifu)."posterior capsule rupture" is indicated as a potential adverse event related to iol implantation in hoya ifu covered under the warnings section.Type of investigation, findings, and conclusion are pending device return and completion of product investigation.Once the product investigation is completed, a follow-up report will be submitted to fda which will include the manufacturer's codes for type of investigation, findings, and conclusion.
|
| |
|
Event or Problem Description
|
|
Damaged haptic after implantation.Lead haptic broke off when unfolding and lead haptic cut the bag.Patient condition: capsular bag tear.Health impact: device explantation.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up #1 emdr is being submitted to fda for a reportable event that occurred in the usa.The report includes corrected information and additional information not available/included in the initial report.Corrected information: h3 - corrected to device not returned.Additional information: g2: added company representative for investigation.G6: type of report - noted as follow-up #1.H2: type of follow-up - noted for corrected and additional information.H6: added codes for manufacturer's investigation: type; findings; and conclusion.The product was not returned to the manufacturer.The investigation was conducted, with the methods and results as noted below.No abnormalities were found in production and inspection records of the product.(serial no: (b)(6); model: b1pc).We also confirmed there were not any abnormalities on the loop pull strength test record of the material lot.(sotj-g115-03).From our investigation, we could not confirm the reported event.Based on available information, we believe this event was not caused by our product quality.A review of the most recent complaint trending data indicates that no significant trends have been identified at this time and no capa is required as part of the product evaluation.
|
| |
|
Event or Problem Description
|
|
Damaged haptic after implantation.Lead haptic broke off when unfolding and lead haptic cut the bag.Patient condition: capsular bag tear.Health impact: device explantation.
|
| |
|