P1008 - clinical outcome - issue: on (b)(6) 2022, dr.(b)(6) (b)(4) reported to (b)(6) via email stating "my patient needed a toriciol, so i used lensar for the cap, 3-plane chop, and intelliaxis.The cap came off easily, and the intelliaxis nubs appeared normal.I did notice that the chop extended pretty deep so the pie pieces separated virtually to the capsule.After removing the 6 pieces of nucleus, i saw a clear lineposteriorly in cortex.I injected viscoat but couldn't tell whether the posterior capsule had ruptured.I proceeded to i/a the cortex and saw ananterior capsule rent emanating from the intelliaxis nub.I then realized that the rent had extended posteriorly.Needless to say, this was not much fun since the rent occurred with an eye full of cortex.But, using the vitrector i removed the cortex but had to bail on the toric lens since the lens needed sulcus fixation.".
|