• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE® TAPERED CERTAIN® IMPLANT 5 X 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE® TAPERED CERTAIN® IMPLANT 5 X 10MM; DENTAL IMPLANT Back to Search Results
Model Number INT510
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Hypersensitivity/Allergic reaction (1907)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant at tooth site #30 was removed due to allergic reaction.Symptoms as a result of the event: inflammation, pain, pressure and swelling.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).Weight unknown / not provided.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: h6: type of investigation codes were updated: 3331, 4109, 4111.H6: investigation findings code was updated: 213.H6: investigation conclusion code was updated: 4310, 50.A osseotiteâ® tapered certainâ® implant 5 x 10mm (int510) was returned for investigation.Visual inspection of the as returned product identified worn markings due to usage, however no damage identified.Based on the evaluation, the device malfunction has not occurred.Additionally, there is no existing nonconformance/capa/hhe/d/ie/product holds against the reported devices that could cause or contribute to the reported event.Zimmer biomet quality management system (qms) has controls in place to ensure the distribution of conforming product.Therefore, based on the available information, the products were likely within specifications and likely conforming when they left zimmer biomet.Dhr review was completed for the subject lot number (2021090385).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Sterilization record (op0210) was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number (2021090385) for similar event and no other complaint was identified.The reported event could not be recreated due to the nature of the dental device and event and the complaint is related to the functional performance of the device.A definitive root cause could not be identified.However, based on the investigation and risk file review, the most likely cause determined from the investigation is clinician places implant in a patient with patient factors (e.G., poor hygiene, periodontal issues, pre-existing medical conditions) and inadequate treatment planning.No further investigation and no immediate capa/hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional event information was received at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® TAPERED CERTAIN® IMPLANT 5 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15083654
Report Number0001038806-2022-01105
Device Sequence Number5161464
Product Code DZE
UDI-Device Identifier00844868006695
UDI-Public(01)00844868006695(17)260924(10)2021090385
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K063341
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date09/24/2026
Device Model NumberINT510
Device Catalogue NumberINT510
Device Lot Number2021090385
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2022
Supplement Date Received by Manufacturer12/20/2022
Initial Report FDA Received Date07/22/2022
Supplement Report FDA Received Date12/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age29 YR
Patient SexFemale
-
-