H11: b2: other serious or important medical events was checked in error on the initial report submitted on august 31, 2022.H10: siemens conducted a detailed investigation of the reported event.The investigation was performed on expert discussions considering the complaint description, customer service reports, system history, system log files, a c-part.The detailed investigation revealed that the cause of the blocked x-ray release was a temporary problem in the internal signal processing.In regular intervals the hardware is checked automatically.Within this check the signal line between relevant sensors and the evaluation component is tested.The test is triggered when a relay switches the contact to the sensors off and back on.The d1-board (evaluation component) detects the falling edge and checks that the voltage drops below a defined "low" level and remains like that for a defined, short period of time.In the present case, it was found that the contact of the relay showed bounce effects/voltage fluctuations.This triggered multiple, successive falling edges that repeated the cable test.However, since the voltage must remain in the "low" state for a certain time for the test to pass, this caused the test to fail even though there was no fault on the signal line.As a result, the x-ray release was prevented to avoid possible severe damage to the x-ray tube.Normal operation of the system was restored by a "power cycle," i.E., the system was shut down and turned back on so that the associated components (x-ray generator, tube) were restored to normal.Field updates (ax028/22/s, ax039/22/s - res 90771) were released for this sporadic problem.The incident described was not classified as a reportable adverse event after detailed investigation, because neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.The corresponding probability according to the risk analysis is in the range of inconceivable.Internal id # (b)(6).
|