| Model Number |
90500 |
| Medical Device Problem Codes |
Overheating of Device (1437); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The led headlight only (90500) was returned for evaluation: the device history record (dhr) was reviewed, and no anomalies related to the reported failure were observed.Failure analysis: evaluation of the led headlight found that the luminaire gets really hot; additionally, the unit had a broken module and rear cover.All parts are to be replaced as per estimate to bring unit back to manufacturer's specifications.Root cause analysis: the returned led headlight was in used condition with damage to the module and rear cover due to rough handling/environmental damage.This most likely lead to the fan working sub-optimally and caused the device to overheat.The reported complaint was confirmed.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
| |
|
Event or Problem Description
|
|
A facility reported that the led headlight only (90500) was overheating and switching itself off.There was no patient involvement and no delay in surgery was reported.
|
| |
|