Model Number 8675324
Medical Device Problem Codes Break (1069); Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Rwmic is submitting this report on behalf of the manufacturer richard wolf gmbh.Device evaluation and inspection was performed by rwmic on 05-july-2022.Device investigation results provided: rwmic investigation report device evaluation, form 5020 (12).The device was tested visually.The reported condition was confirmed and the device did not meet specifications.The evaluation found that the damage is consistent with an impact against a hard surface.Damage may have caused a thin fracture that help together until there was stress on the device during use of breaking apart immediately.The device was repaired.Additional information regarding device labeling included.(b)(4), section 8.1.2."shark" resectoscope, pg 27.Warning! danger of injury! incorrect handling, e.G.Fall, shock, blows or similar mechanical loads can cause hair cracks and / or spalling of the ceramic coating in the distal area of the resectoscope sheath.Injuries of the patient, user or third parties are possible.Mind surface changes and ensure safe handling.Do not use damaged resectoscope sheaths, return damaged sheaths for repair.Check the ceramic insulation at the distal end of the resectoscope sheath for damage before every use.Replace damaged or brittle o-rings / caps (12.13).O-rings (1.3) (1.5) - "shark" resectoscope, rotatable [a ] [ b ] o-ring (7.5) - "shark" resectoscope with central stopcock [ c ] non-rotatable rw gmbh considers this mdr open.Rw gmbh will submit a follow up report when new information becomes available.
 
Event or Problem Description
Richard wolf medical instruments corp.(rwmic) reference (b)(4).It was reported to richard wolf medical instruments corp.(rwmic) that "tip broke off during a procedure.Delay time was 15 minutes." additional information reported by the initial reporter: will the device be returned? yes.Was the device being used on a patient when the reporting issue occurred? yes.Was there any injury or illness to the patient due to the reported issue? no.Was there any injury or illness to any other personnel due to the reported issue? no.Did the issue cause a delay in the procedure being performed? yes.Did the delay put the patient at risk? no.Was there a similar back-up device available for use? yes.Was the scheduled procedure completed? yes.Additional details about the adverse event; (b)(6) 2022.(1) what was the scheduled procedure being performed when the tip broke? transurethral incision of the prostate (tuip).(2) did the tip break inside or outside of the patient? yes.(3) what actions were being done during the delay? remove the tip of the sheath from the patient.(4) did the patient require any additional treatment during the delay or will require another procedure as the result of the delay? no.(5) what was the patient's outcome? no problem at the time of discharge.Rwmic mdr awareness date: (b)(6) 2022 also, the purpose of this submission is to provide the results of the device investigation.
 
Additional Manufacturer Narrative
Rwmic is submitting this report on behalf of the manufacturer richard wolf gmbh.Follow-up report #1 is to provide fda with new information received from the user facility.Rw gmbh considers this mdr closed.Should rwgmbh receive new information, a follow-up report will be submitted.
 
Event or Problem Description
Richard wolf medical instruments corp.(rwmic) reference complaint: (b)(4).The purpose of this report is to update mdr with new patient information, received from a representative of the user facility on 8/16/2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHARK
Common Device Name
INTERNAL SHEATH RESECTOSCOPE 24FR
Manufacturer (Section D)
RICHARD WOLF GMBH
pforzheimer strasse 32
75438 knittlingen, germany
GM 
Manufacturer (Section G)
RICHARD WOLF GMBH
pforzheimer strasse 32
d-75438 knittlingen, germany
GM  
Manufacturer Contact
zsófia bartholome-nagy
pforzheimer strasse 32
d-75438 knittlingen, germany 
GM  
MDR Report Key15084780
Report Number9611102-2022-00033
Device Sequence Number4055583
Product Code FJL
UDI-Device Identifier04055207019658
UDI-Public04055207019658
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K062720
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Model Number8675324
Device Catalogue Number8675324
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2022
Supplement Date Received by Manufacturer08/16/2022
Initial Report FDA Received Date07/22/2022
Supplement Report FDA Received Date08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/17/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age81 YR
Patient SexMale
Patient Weight67 KG
Patient RaceWhite
-
-