SYNTHES GMBH 2.7MM TI METAPHYSEAL SCR/SLF- TPNG W/T8 STRDRV RECESS/18MM; SCREW, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
04.118.518 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The initial reporter is j&j company representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2022, the patient underwent an open reduction with internal fixation for fracture of shaft of clavicle.When the surgeon inserted the 2nd screw to a hole of a plate, the screw broke at its neck.The surgeon removed the plate, tried to extract the screw shaft left in the bone with radio pliers, etc., but he couldn¿t.The sales rep proposed to extract it by shaving the bone around the screw shaft with a canulated drill, and the surgeon could finally extract the screw shaft.Since the extraction of the screw shaft made a big hole on the bone, the surgeon tried to use another plate, but the plate did not fit to the bone due to shape mismatch with the bone.The surgeon decided to use the original plate for the treatment.The surgery was completed successfully with 90 minutes delay.It was noted the patient had hard bones.There are no pieces in the patient.The surgeon confirmed by x-ray.Patient is listed as stable.This is report 1 of 1 for (b)(4).This report is for a screw.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d6a.D9.H3, h4, h6: part: 04.118.518ts.Lot: 8l72827.Release to warehouse date: 26.Oct.2021.Expiration date : 01.Oct.2026.Supplier: (b)(4).Manufacturing site: werk selzach.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that lwprof metaphys compres scr ã¸2.7 self-ta was observed to be broken under screw head.The broken head was received in the evidence provided.A dimensional inspection for the lwprof metaphys compres scr ã¸2.7 self-ta was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the lwprof metaphys compres scr ã¸2.7 self-ta would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|