MAUDE Adverse Event Report: COOK INC COOK® SINGLE-USE HOLMIUM LASER FIBER; GEX LASER INSTRUMENT, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC COOK® SINGLE-USE HOLMIUM LASER FIBER; GEX LASER INSTRUMENT, SURGICAL Back to Search Results
Model Number G25294
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The user facility medwatch report # mw5110034 was received and the report showed that the customer reported: during a ureteroscopy with laser lithotripsy for left urolithiasis, a 2 mm piece of a cook® single-use holmium laser fiber broke during procedure and could not be retrieved from the patient's kidney.Attempts to find the piece were abandoned, as the risks of continuing to search for it did not outweigh the benefits.There is low risk to the patient especially since the piece of fiber may wash out on its own.A clinical disclosure was done with the patient and his family, and this was documented in the operative report.The patient will have follow-up which includes a computed tomography (ct) scan at about 4 weeks post removal of the ureteral stent that was placed during the procedure.No additional patient consequences were reported.Additional information has been requested.At the time of this report, no further information has been provided.
 
Additional Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Additional Manufacturer Narrative
Summary of event the user facility filed a medwatch report [mw5110034] on (b)(6) 2022.This report was received by cook on 27jun2022.The report stated that a single-use holmium laser fiber [g25294-hlf-s273-hsma; 14120540] was used for a ureteroscopy with laser lithotripsy for left urolithiasis on (b)(6) 2022.No additional information was provided regarding patient pre-existing conditions, anatomy, etc.It was reported that during the procedure a 2mm segment of the device broke and could not be retrieved from the patient¿s kidney.Attempts to find the piece during the procedure were abandoned, as the user felt the risks of continuing to search for it did not outweigh the benefits.A clinical disclosure was done with the patient and his family, and this was documented in the operative report.The patient will have follow-up which includes a computed tomography (ct) scan at about 4 weeks post removal of the ureteral stent that was placed during the procedure.Investigation evaluation reviews of the complaint history, device history record, instructions for use, manufacturer¿s instructions, and quality control procedures were conducted during the investigation.The complaint device was not returned to cook for investigation.Due diligence was completed to the retrieve the complaint device, however the product has yet to be received.If the product is received at a later time, the complaint file will be updated accordingly.Accordingly, no physical examination could be performed.A document-based investigation evaluation was performed.A review of the device history record found no related nonconformances.A complaint history database search revealed no other complaints associated with lot number 14120540.Because there were no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in the field.The device is provided with instructions for use which caution, "a number of different factors affect the life of any particular fiber, including: continuous lasing with the fiber tip in contact with tissue.Never subject fiberoptics to sharp bends in handling, use, or storage." based upon the available information and results of the investigation, cook has concluded that the cause for the complaint was due to an unintended user error that caused or contributed to event.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Correction of date of event from 25jun2022 to 25may2022 as per medwatch report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK® SINGLE-USE HOLMIUM LASER FIBER
Common Device Name
GEX LASER INSTRUMENT, SURGICAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15086315
Report Number1820334-2022-01261
Device Sequence Number17433561
Product Code GEX
UDI-Device Identifier00827002252943
UDI-Public(01)00827002252943(17)240730(10)14120540
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K133788
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/30/2024
Device Model NumberG25294
Device Catalogue NumberHLF-S273-HSMA
Device Lot Number14120540
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/27/2022
Supplement Date Received by Manufacturer10/10/2022
Initial Report FDA Received Date07/22/2022
Supplement Report FDA Received Date11/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-