| Model Number |
M00535150 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the device analysis, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a stonetome was used in the bile duct during an endoscopic bile duct stone removal procedure performed on (b)(6) 2022.During the procedure, the tip of the forming mandrel was detached from the device and fell into the patient.The detached tip was retrieved using a forceps.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.Note: a photo of the complaint device outside the patient was provided by the customer and shows the detached tip of the forming mandrel.No further information has been obtained despite good faith efforts.
|
| |
|
Additional Manufacturer Narrative
|
|
Impact code f2301 captures the reportable event of retrieval of the detached forming mandrel tip using a forceps.Medical device problem code a0203 captures the reportable event of forming mandrel tip detached.The returned curving mandrel of the stonetome was analyzed, and a visual evaluation noted that the molding tip was detached from the curving mandrel and it was not returned.Per media analysis, the provided photo showed that the molding tip was detached from the curving mandrel.The curving mandrel was observed under magnification and no visual damage/problem found.Per the dimensional inspection, the thickness and the length of the coined end of the were measured and these were within specifications.No other problems with the device were noted.The reported event of molding tip was detached from the curving mandrel was confirmed.Upon analysis, it was found that the molding tip was detached from the curving mandrel which matches with the findings based on the media inspection; however, it was not returned.Based on the findings, the problem could have been generated due to handling/manipulation upon removal of the mandrel from the stonetome.Therefore, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a stonetome was used in the bile duct during an endoscopic bile duct stone removal procedure performed on (b)(6) 2022.During the procedure, the tip of the forming mandrel was detached from the device and fell into the patient.The detached tip was retrieved using a forceps.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.Note: a photo of the complaint device outside the patient was provided by the customer and shows the detached tip of the forming mandrel.No further information has been obtained despite good faith efforts.
|
| |
|