EXACTECH, INC. TRULIANT TIB FIT TRAY CEM SZ 3F / 2T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-022-45-3020 |
| Medical Device Problem Codes |
Loss of Osseointegration (2408); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Synovitis (2094); Aseptic loosening (4883)
|
| Date of Event |
06/24/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this female patient's tka was revised due to pain and synovitis.Tibial component was found loose.Patient left in stable condition with a new tibia implanted with stems and augments, and new 13mm psc insert.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (d4) catalog number: 02-022-45-3020, serial number: (b)(6), expiration date: 16-jun-2029, unique identifier (udi) #: (b)(4).(d10) concomitant device(s): a10012, (b)(6) - gps implant kit v2.02-022-44-3013, (b)(6) - truliant tib imp psc insert sz 3, 13mm.02-020-11-0230, (b)(6) - truliant ps cem fem ps cem left sz 3.200-02-35, (b)(6) - three peg patella 35mm.201-78-15, (b)(6) - holding pin mini sharp point 4 pk.521-78-23, (b)(6) - threaded pin size 2.3 collared 2pk.521-78-36, (b)(6) - threaded pin size 5.1 collarless 2pk.521-78-32, (b)(6) - threaded pin size 3.0 collarless 2pk.521-78-32, (b)(6) - threaded pin size 3.0 collarless 2pk.(e1) first name: (b)(6).(h3) the revision reported may have been the result of an insufficient bond between the implant and the bone which led to aseptic (non-infected) loosening of the tibial tray and subsequent pain.However, this cannot be confirmed because the component was not returned for evaluation.(h4) device manufacture date: 18-jun-2019 section h11: *the following sections have corrected information: (d1) brand name: truliant tib fit tray cem sz 3f / 2t.(h6) component code: 4755, part/component/sub-assembly term not applicable.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were added/corrected: b1, b2, d6a, d8, g1, h6: clinical code, medical device problem code, component code, investigation findings, investigation conclusion.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|