| Model Number |
01-003-02-0003 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the cup impactor broke.
|
| |
|
Additional Manufacturer Narrative
|
|
The broken device reported was likely the result of numerous impaction forces from repeat surgical use, which led to crack initiation, propagation, and ultimate failure of the body-strike plate weld interface.However, this cannot be confirmed as the device was not returned for evaluation.The following sections have corrected information: component code: 3195, weld.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment of the reported event, it was determined that the event did not involve a death or serious injury nor would be likely to cause or contribute to a death or serious injury if the reported malfunction were to recur.As a result, this complaint event is no longer considered reportable by exactech and this report may be disregarded.
|
| |
|