• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACCURAY INCORPORATED CYBERKNIFE TREATMENT DELIVERY SYSTEM; MEDICAL LINEAR ACCELERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACCURAY INCORPORATED CYBERKNIFE TREATMENT DELIVERY SYSTEM; MEDICAL LINEAR ACCELERATOR Back to Search Results
Model Number G4
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A mounting nut on the synchrony boom arm was loose due to the fse not retightening the nut enough.Investigation is in-progress.
 
Event or Problem Description
The synchrony boom arm had unintended motion.No death or serious injury was reported.
 
Additional Manufacturer Narrative
The field service engineer (fse) went to service the system and found an issue that would not allow the completion of the service event.After the fse re-assembled the system it was discovered the boom arm screw as loose.The nut became loose because the fse did not tighten it enough when he worked on the device.The system has since been corrected and the loose hardware was re-tightened.No further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYBERKNIFE TREATMENT DELIVERY SYSTEM
Common Device Name
MEDICAL LINEAR ACCELERATOR
Manufacturer (Section D)
ACCURAY INCORPORATED
1209 deming way
madison WI 53717
Manufacturer (Section G)
ACCURAY INCORPORATED
1209 deming way
madison WI 53717
Manufacturer Contact
daniel biank
1209 deming way
madison, WI 53717
6088242971
MDR Report Key15089213
Report Number3003873069-2022-00012
Device Sequence Number5101286
Product Code IYE
UDI-Device IdentifierM6580325000
UDI-Public+M6580325000/16D20190227E
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K170788
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG4
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2022
Supplement Date Received by Manufacturer03/29/2022
Initial Report FDA Received Date07/22/2022
Supplement Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/27/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-