The customer reported that during a therapeutic sphincterotomy procedure, the knife wire broke when it was bowed and cautery was applied.The device was half-bowed at the time, and it occurred during the third time that cautery was applied.The intended procedure was successfully completed.There was no patient or user injury reported due to the event.Additional details have been requested regarding the reported issue.At this time, no additional information has been provided.
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.Olympus performed a visual inspection on the device and noted some abnormalities where the cutting wire was broken and detached on one end.The coated portion of the cutting wire was still attached to the cutting wire however it appeared slightly damaged.The portion of the cutting wire located near the "green" distal marker on the distal end has signs of burnt char marks.The cutting wire itself was severely damaged and deformed as there appear to be minor kinks on the cutting wire.The handle portion appeared to have no signs of physical damage on the slider, injection port, and plug.An olympus evaluator gently ran their fingertips over the entire length of the insertion portion and was unable to identify any crushed, broken, or kinked areas.Functional tests were unable to be carried out as the cutting wire was severely damaged and detached.Pictures of the device revealed the following: ·the cutting wire was broken on the coated portion.·the broken portion of the cutting wire was scorched.·the coated portion of the cutting wire was torn.·the lot number was 17v with supplementary information number of ¿07¿.Therefore, the device was manufactured on july 7, 2021.·the model number was kd-v611m-07301s.However, it could not be determined from the pictures that the cutting wire had fallen off.A review of the device history record found no deviations that could have caused or contributed to the reported issue.For the following reasons, it was determined that the reported event was not the cutting wire falling off, but the breakage of the cutting wire.The event description did not contain that the cutting wire had fallen off ·it was not confirmed by the device evaluation that the cutting wire had fallen off.Based on the investigation result and the result of past similar complaint investigation, a likely mechanism causing the cutting wire breakage might be the following.However, the exact cause could not be determined.1.The cutting wire where the wire coating was torn came into contact with the distal end of the endoscope when the forceps elevator was raised.2.Under circumstances described above 1, an electric conduction was activated.This caused the cutting wire to become hot instantly at the contact point.As a result, the cutting wire broke.Also, it is confirmed that the tear of the coated portion of the cutting wire can replicated by the following mechanism: 1.Raise the forceps elevator of the endoscope.2.Once the cutting wire deflects, the coated portion of the wire comes into contact with the metal part of the distal end of the endoscope.3.The cutting wire was moved back and forth under the circumstances described above 2, causing the coated portion to tear.It is possible that the slider was slightly pushed hard.This possibly caused the cutting wire to deflect.The following is included in the instructions for use (ifu): "¿since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.¿when inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.¿do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.¿if you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion." olympus will continue to monitor field performance for this device.
|