Model Number 383592
Medical Device Problem Code Material Discolored (1170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd nexiva¿ diffusics¿ closed iv catheter system experienced discoloration.The following information was provided by the initial reporter: catheter, tip is discolored.
 
Additional Manufacturer Narrative
(b)(6).A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
G5: corrected 510k number k123734 to k173354 h6: investigation summary our quality engineer inspected the 2 representative samples submitted for evaluation.The reported issue of discoloration was confirmed upon inspection of the samples.Both samples were visually inspected, and one of the two samples showed discoloration.Bd determined that the cause of the discoloration can be associated to our laser cutting process during manufacturing.The discoloration is slight burning of the material due to the heat from the cutting laser.However, the amount of discoloration found was acceptable based off our manufacturing standards.Complaints received for this device and reported condition will continue to be tracked and trended.
 
Event or Problem Description
It was reported that the bd nexiva¿ diffusics¿ closed iv catheter system experienced discoloration.The following information was provided by the initial reporter: catheter, tip is discolored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD NEXIVA¿ DIFFUSICS¿ CLOSED IV CATHETER SYSTEM
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15090052
Report Number9610847-2022-00271
Device Sequence Number14742764
Product Code FOZ
UDI-Device Identifier30382903835929
UDI-Public30382903835929
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K173354
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/31/2025
Device Model Number383592
Device Catalogue Number383592
Device Lot Number2090066
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2022
Supplement Date Received by Manufacturer08/09/2022
Initial Report FDA Received Date07/22/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-