It was reported that, after a right tha surgery was performed on (b)(6) 2011, the patient experienced a polarstem stem lat.Ti/ha 4 non-cem fracture after a fall.This adverse event was addressed via revision surgery on (b)(6) 2022, where the femoral stem and head were explanted.Current health status of patient is unknown.
|
|
H3, h6: it was reported that, after a right total hip arthroplasty surgery was performed on (b)(6) 2011, the patient experienced a polarstem neck fracture after a fall.The broken polarstem with the ball head, used in treatment, were received for investigation.The relationship between the reported event and the device can be confirmed, the neck of the polarstem is fractured.Furthermore, there are a lot of scratches visible on the surface of the stem, most likely coming from the reported revision surgery.A material assessment was conducted.About two thirds of the fracture surface show fatigue striations, indicating that the crack propagated by fatigue.The residual fracture surface is covered by dimples, characteristic of ductile overload.No material defects, which could have initiated or enhanced crack growth, could be observed on the fracture surface.In the area of assumed crack initiation, scratches, stainless steel residues and discolorations from the potential use of an electric scalpel have been found which were likely present prior to fracture and may have led to crack initiation.Since the batch number is covered by the ball head, a review of the production documentation cannot be conducted without disassembling the devices.Therefore, it cannot be concluded that the part match manufacturing specification at the time of release.A complaint history review showed that the complaint is in line with the current risk management.The risk level is low and covered through the risk management files.The reported failure mode is listed among the potential medical device problems.A review of past corrective actions was performed.No further escalation is required.A medical assessment was performed based on the provided implantation report.Other medical documentation was not received for investigation.Based on only the implantation report a thorough medical assessment cannot be conducted.However, the reported patient's fall cannot be ruled out as a contributing factor to the polarstem fracture.Furthermore, it is reported in the implantation report that a ceramic ball head was implanted.However, the returned ball head is a metal ball head.This indicates that a ball head revision surgery took place between the implantation and the revision surgery due to the fracture of the stem.It is therefore possible that the neck of the polarstem was scratched or damaged by the use of an electric scalpel, which could have led to the crack initiation as described above.The root cause can not be determined after investigation.It cannot be excluded that the scratches and damages on the neck, which might have happened during the ball head revision, contributed to the fatigue fracture.To date, further actions are not deemed necessary.Smith and nephew will monitor this device for further similar issues.Both devices will be retained.Internal complaint reference number: (b)(4).
|