(b)(4).Device passed the functional test, including the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and the delivery accuracy test at 0.08715 inches.The insulin flow blocked alarm functions properly during the basic occlusion test, occlusion test and force sensor test.No unexpected insulin flow blocked alarm/no delivery alarm noted during testing.No keypad backlight anomaly noted during testing.Device communicated properly with the guardian link 3 and the test plug.Device was able to enter into auto mode properly.No insulin pump/sensor communication anomaly, lost sensor alarm, lost sensor signal alert, or calibration anomaly noted.Insulin pump had cracked keypad overlay at the select button, a pillowing keypad overlay, scratched case and a cracked retainer.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.The type of investigation, investigation findings and investigation conclusion codes that were provided with the initial report were incomplete.The complete information has been included with this report in h6 section.
|