|
The exact date of event is unknown, therefore, the date of the intervention was used as date of event.Concomitant medical products: gore® excluder® aaa endoprosthesis - stent graft contralateral leg endoprosthesis: pxc141400 sn(b)(4), pxc141200 sn(b)(4), and pxc181000 sn(b)(4).Gore® excluder® aaa endoprosthesis - stent graft iliac extender endoprosthesis.Pxl161407 sn(b)(4).(b)(4).W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
The following literature article was reviewed: bonardelli, s., verzini, f., rivolta, n., pagliariccio, g., zanotti, c., boero, m., franchin, m., carbonari, l., baggi, p., gibello, l., parlani, g., cavi, r.And piffaretti, g.(2022).'long-term outcomes of endovascular aortic repair with flared iliac limb endografts in patients with abdominal aortic aneurysm and aneurysmal common iliac arteries', j cardiovasc surg (torino), 63(4), pp.464-470.The aim of this study was to evaluate mid-to-long-term outcomes of endovascular aneurysm repair using flared iliac limb grafts in patients with abdominal aortic aneurysm (aaa) and aneurysmal common iliac arteries (cias).This is a multicenter, retrospective, observational cohort study that involves four tertiary referral hospitals between may 1, 2005 and april 30, 2019.995 aneurysmal iliac limbs in 795 patients who met the inclusion criteria were involved in the study.Out of these 795 patients, 471 patients were treated with zenith alpha¿ (cook), 180 patients with endurant¿ (medtronic), and 144 patients were treated with gore® excluder® aaa endoprosthesis.It was additionally reported by the corresponding author, that a total of 90 patients was treated in university of insubria school of medicine, were 55 patients underwent endovascular aneurysm repair (evar) using gore® excluder® aaa endoprosthesis.One patients underwent reintervention due to explant.The initial evar procedure on december 28, 2010 was uneventful.Due to a distal migration (>10mm) of the device at proximal neck with sac enlargement and coexistent persistent type 2 endoleak from the lumbar arteries, the device was explant on april 9, 2019 and the procedure was uneventful.
|
|
A review of the manufacturing records indicated the lot met pre-release manufacturing specifications.Engineering evaluation could not be conducted as the devices are not returned.B13: the initial reporter (the corresponding author or the literature article) has been contacted and event dates, event details, as well as patient specific information was received.Per gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and/or require intervention include, but are not limited to: component migration, endoleak.Another event within this article is reported with gore mpdcase-00009742-1 number (manufacturer report number 2017233-2022-03150).Corrected h6: updated health effect - clinical code to e050102 and e2121.Updated investigation findings code to c19 - no device problem found.Code 3233 is no longer applicable.Updated investigation conclusions code to d15 and d12.Code 11 is no longer applicable.
|