• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA MUST MINI CANNULATED POLYAXIAL SCREW 4.0 X 30 FULL THREAD + NUT; SPINE CANNULATED SCREW'S SET SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA MUST MINI CANNULATED POLYAXIAL SCREW 4.0 X 30 FULL THREAD + NUT; SPINE CANNULATED SCREW'S SET SCREW Back to Search Results
Model Number 03.75.210
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 05 july 2022: lot 2020760: (b)(4) items manufactured and released on 28-may-2020.Expiration date: 2025-05-12.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event during the period of review.
 
Event or Problem Description
At about 5 months after the primary, it was observed that the set screw inserted to the c2 screw came off.The surgeon revised the screw in c2.And the surgery was completed successfully.
 
Additional Manufacturer Narrative
Visual inspection performed by medacta spine r&d project manager: during the visual inspection on retrieved implants it was observed that the set-screw were still intact with no sign of threading deformation or breakage.A sawbone was performed on the retrieved set-screw component in order to verify the overall functionality when secured with the associated polyaxial screw component.During the test it was observed a slight friction at the threading interface during advancing.However it was possible to position and secure the set-screws as per standard procedure under 3nm torque.Reverse procedure was successfully performed to remove the set-screw all the way.The test was successfully performed three time with no evidence of implants failure.Based on test result, it was possible to secure and remove the set-screw definably, according to the standard procedure and available instrumentation.The reason of loosening is unknown.It is supposed that the set-screw was not fully secured during the final procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUST MINI CANNULATED POLYAXIAL SCREW 4.0 X 30 FULL THREAD + NUT
Common Device Name
SPINE CANNULATED SCREW'S SET SCREW
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15092005
Report Number3005180920-2022-00567
Device Sequence Number17346883
Product Code NKG
UDI-Device Identifier07630030863882
UDI-Public07630030863882
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K171369
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/12/2025
Device Model Number03.75.210
Device Catalogue Number03.75.210
Device Lot Number2020760
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/05/2022
Supplement Date Received by Manufacturer07/05/2022
Initial Report FDA Received Date07/25/2022
Supplement Report FDA Received Date11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-