(b)(4).Pump was received with a cracked battery tube threads and a cracked case behind the pump near the battery tube compartment.The test reservoir does lock properly inside the reservoir tube.Pump passed the displacement test and self test.The pump was programmed with a test guardian link (3) transmitter and a 240 mg/dl glucose sensor simulator.The pump communicated properly with glucose sensor simulator and displayed the calibrate your sensor alarm properly after completion of the warm up.The pump calibrated and displayed the programmed value of 239 mg/dl properly on the display graph.No unexpected sensor errors or anomalies were noted during testing.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|