| Model Number |
621-400ES |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the user facility: primary nurse (rn) used dose mode to program fentanyl with co-signing rn as primary medication.Red screen on left read "fentanyl" as drug name.Patient transported to ct scan via rn and renal therapist.Rn received phone call around noon from icu provider that family had decided to transition to comfort measures only and ct scan canceled.Approximately twenty minutes later arrived back to unit with patient.30 minutes later, dose increased to 37.5 mcg/hr.An hour later rn entered room due to iv pump alarming infusion complete (fentanyl).Found fentanyl running at 37.5ml/hr, per medication administration record (mar) should have been running at 37.5 mcg/hr for a rate of 1.9ml/hr.Icu provider informed of above.Cosigning rn aware of above.Pump immediately sequestered.Fentanyl infusion placed on new pump with new bag.
|
| |
|
Additional Manufacturer Narrative
|
|
Thie report has been identified as b.Braun medical, inc.Internal report (b)(4).The complaint is under evaluation.A follow-up report will be provided after the examination results are available.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical, inc.Internal report (b)(4).The device was not returned for evaluation.Further investigation of the complaint is not possible without a device.If the device does become available, the complaint will be reopened for further evaluation.All information concerning this reported incident has been included in our trend analysis of the product line.
|
| |
|
Search Alerts/Recalls
|