Model Number CNA0T0
Medical Device Problem Codes Break (1069); Failure to Eject (4010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A pharmacist reported that during implantation of intraocular lens (iol), while injecting the lens stuck in the incision site.The haptics broke when removing the lens.The surgery was completed with the same power and model back up lens.There was no patient harm.
 
Additional Manufacturer Narrative
The product was returned for analysis and the reported complaint was observed.Iol returned in a specimen container in an unknown, clear liquid.The iol is air dried.Solution is observed on the iol.One haptic is broken/torn and is returned.The optic is torn/cracked/ nicked/ chipped/ scratched/marked.The product investigation could not identify the root cause for the reported complaint.The device was not returned.Only the iol returned.A definitive determination of the possible root causes cannot be made without the evaluation of the involved device.A final root cause cannot be determined based on available information.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15092776
Report Number9612169-2022-00350
Device Sequence Number14656134
Product Code HQL
UDI-Device Identifier00380652393591
UDI-Public00380652393591
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 10/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/14/2024
Device Model NumberCNA0T0
Device Lot Number25140656
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/27/2022
Supplement Date Received by Manufacturer09/15/2022
Initial Report FDA Received Date07/25/2022
Supplement Report FDA Received Date10/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC OVD.
Patient Age74 YR
Patient SexUnknown
-
-