| Model Number |
GELWEAVE VALSALVA |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Pericardial Effusion (3271)
|
| Date of Event |
06/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Type of investigation codes: analysis of production records - a review of the retained manufacturing/quality records confirmed that the device was manufactured to specifications.No previous complaints/issues have been logged for the remainder of the batch.Specific emphasis was directed toward the porosity results; which clearly indicate good results in the lower end of the test.Historical event analysis & trend analysis: a review of similar events for gelweave branded devices provided an occurrence (b)(4) (complaint v sales).No reports of leakage has been received for any of the remaining grafts from this batch.No negative trend was identified.Communication / interviews.- further information was requested from the site with regards to the circumstances of the event.Device not accessible for testing - device remains in situ and was not explanted.Investigation findings: no device problem found.Review of all retained manufacturing records (including porosity test records) confirm that the device was manufactured to specifications.Investigation conclusion: cause not established.Based on all available information provided by the site and that the actual device is not available, no determination of a root cause has been possible.Further action is not planned, however, the issue will be tracked and trended as part of the on-going complaints trending and reporting process and if an adverse trend develops action may be taken at that time.
|
| |
|
Event or Problem Description
|
|
This event was reported via the panther study as follows: a gelweave valsalva device was implanted on (b)(6) 2022.On (b)(6) 2022 it was detected that the patient was suffering from pericardial effusion (build-up of fluid within the pericardium sac); which was treated with a re-intervention by suturing bleeding at the distal anastomosis.
|
| |
|
Search Alerts/Recalls
|