(b)(4).Concomitant medical products: 47-2486-050-40 3068483 blunt tip screw 4x50mm; 47-2488-000-04 3008331 ann washer small; 47-2486-052-40 3024745 blunt tip screw 4x52mm; 47-2496-260-07 3008140 ann proximal humerus, right 7x260mm.Source: foreign ¿belgium.Multiple mdr reports were filed for this event, please see associated reports: 0009613350 - 2022 - 00382, 0009613350 - 2022 - 00384 and 0009613350 - 2022 - 00387.An investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
|
The following sections were updated: b4, b5, d9, g3, g6, h1, h2, h3, h6, h10.Two of the complained blunt tip screws (52mm and 50mm) as well as a washer were returned for examination.Multiple scratches and signs of wear can be seen on both the screws and the washer.Additionally, the 52mm blunt tip screw shows signs of thread stripping, and the 50mm blunt tip screw has damaged threads.Both of these phenomena are indications of screw migration.Review of the device history records identified no deviations or anomalies during manufacturing.Devices are used for treatment.Based on the investigation it could be assumed that further possible contributing factors to the migration of the screw might be multifactorial related to either patient condition, behaviour or implantation procedure.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.The investigation also considered whether the design limitation of the corelock mechanism could be a potential contributing factor.However, further biomechanical testing identified that the product performance is in an acceptable range in comparison with legally marketed similar devices.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.
|