| Model Number |
6197-9-001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The following was reported: "the ampoule bottle breaks every time i open the monomer.It's dangerous and i want the package to be changed.".
|
| |
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding crack/fracture involving a simplex ampoule was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there has been one other similar event for the lot referenced for the same device within the same pi, therefore no further trending is required for this commonality.Conclusions: the event was not confirmed as insufficient information was provided.Further information such as images/video of the event as it occurred as well as return of the device and primary operative reports are required to complete the investigation and determine a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
The following was reported: "the ampoule bottle breaks every time i open the monomer.It's dangerous and i want the package to be changed.".
|
| |
|
Search Alerts/Recalls
|