• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BIO-COMP SWIVELOCK C, CLD 5.5X19.1MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BIO-COMP SWIVELOCK C, CLD 5.5X19.1MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number BIO-COMP SWIVELOCK C, CLD 5.5X19.1MM
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a shoulder arthroscopy surgery the screw broke when the surgeon put the screw on the blue screwdriver.The surgeon turned it and it broke off prior to full insertion to intended depth inside the patient.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Complaint confirmed.Upon visual inspection, it was noted that the sutures attached to the suture eyelet had been removed from the shaft and the device was reassembled to the driver with the sutures outside of the shaft.It was also noted that the screw was broken.The cause of this remains undetermined, however, probable cause can be attributed to prying/leveraging forces being used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-COMP SWIVELOCK C, CLD 5.5X19.1MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15093392
Report Number1220246-2022-05284
Device Sequence Number5161305
Product Code MAI
UDI-Device Identifier00888867097919
UDI-Public00888867097919
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K203495
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2024
Device Model NumberBIO-COMP SWIVELOCK C, CLD 5.5X19.1MM
Device Catalogue NumberAR-2323BCC-1
Device Lot Number12320724
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2022
Supplement Date Received by Manufacturer07/06/2022
Initial Report FDA Received Date07/25/2022
Supplement Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-