| Model Number |
24703 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 70% stenosed target lesion was located in the superficial femoral artery.A 2.5mm x 120mm x 150cm sterling sl and 2.5mm x 80mm x 150cm coyote balloon catheter were advanced for dilation.Both balloon reached the lesion site.However, during inflation at 12 atmospheres, the balloon ruptured.The procedure was completed with another of same device.There were no patient complications reported.The patient was in stable condition following the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter facility name: (b)(6) hospital.Device evaluated by mfr.: returned product consisted of a sterling sl balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed pinhole 64mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 70% stenosed target lesion was located in the superficial femoral artery.A 2.5mm x 120mm x 150cm sterling sl and 2.5mm x 80mm x 150cm coyote balloon catheter were advanced for dilation.Both balloon reached the lesion site.However, during inflation at 12 atmospheres, the balloon ruptured.The procedure was completed with another of same device.There were no patient complications reported.The patient was in stable condition following the procedure.
|
| |
|
Search Alerts/Recalls
|