Batch review performed on (b)(6) 2022, lot 189374: lot 189374: 20 items manufactured and released on 25-apr-2019.Expiration date: 2024-feb-19.No anomalies found related to the problem.To date, all items of the same lot have been sold with 1 similar reported event during the period of review.Additional implanted components: batch review performed on 11 july 2022 gmk-hinge 02.09.0220uch uc fixed tibial insert size 2/20mm (k172347) lot.183630: lot 183630: 11 items manufactured and released on 21-aug-2018.Expiration date: 2023-jul.24.No anomalies found related to the problem.To date, 6 items of the same lot have been sold with 3 similar reported events during the period of review.Batch review performed on 11 july 2022 gmk-hinge 02.09.4002r fixed tibial tray size 2 r (k130299) lot.189378: lot 189378: 24 items manufactured and released on 14-dec-2018.Expiration date: 2023-dec-04.No anomalies found related to the problem.To date, 23 items of the same lot have been sold with no similar reported event during the period of review.
|
|
The patient came in due to signs of an infection and the pathogen is unknown.At 3 years and 2 months after the primary surgery, the surgeon performed a washout, removed all components, and fused the patient's knee.The surgery was completed successfully.
|