• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA GMK-HINGE FEMORAL COMPONENT SIZE 3 R; KNEE FEMORAL COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA GMK-HINGE FEMORAL COMPONENT SIZE 3 R; KNEE FEMORAL COMPONENT Back to Search Results
Model Number 02.09.2603R
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on (b)(6) 2022, lot 189374: lot 189374: 20 items manufactured and released on 25-apr-2019.Expiration date: 2024-feb-19.No anomalies found related to the problem.To date, all items of the same lot have been sold with 1 similar reported event during the period of review.Additional implanted components: batch review performed on 11 july 2022 gmk-hinge 02.09.0220uch uc fixed tibial insert size 2/20mm (k172347) lot.183630: lot 183630: 11 items manufactured and released on 21-aug-2018.Expiration date: 2023-jul.24.No anomalies found related to the problem.To date, 6 items of the same lot have been sold with 3 similar reported events during the period of review.Batch review performed on 11 july 2022 gmk-hinge 02.09.4002r fixed tibial tray size 2 r (k130299) lot.189378: lot 189378: 24 items manufactured and released on 14-dec-2018.Expiration date: 2023-dec-04.No anomalies found related to the problem.To date, 23 items of the same lot have been sold with no similar reported event during the period of review.
 
Event or Problem Description
The patient came in due to signs of an infection and the pathogen is unknown.At 3 years and 2 months after the primary surgery, the surgeon performed a washout, removed all components, and fused the patient's knee.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-HINGE FEMORAL COMPONENT SIZE 3 R
Common Device Name
KNEE FEMORAL COMPONENT
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15093711
Report Number3005180920-2022-00571
Device Sequence Number4034488
Product Code KRO
UDI-Device Identifier07630030825446
UDI-Public07630030825446
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130299
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/19/2024
Device Model Number02.09.2603R
Device Catalogue Number02.09.2603R
Device Lot Number189374
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/05/2022
Initial Report FDA Received Date07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/25/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-