• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH TOTAL BILIRUBIN; DIAZO COLORIMETRY, BILIRUBIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH TOTAL BILIRUBIN; DIAZO COLORIMETRY, BILIRUBIN Back to Search Results
Model Number 6L45-22
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed falsely decreased architect total bilirubin results after installing a new reagent cartridge (same lot number but new cartridge) on (b)(6) 2022.The customer uses a normal range of 0.2 - 1.2 mg/dl.The customer stated they recalibrated the assay using the same reagent cartridge, then repeated all the patient samples which recovered higher.The customer provided 4 examples.Initial low results were reported to the physician.Sid (b)(6): 0.23 mg/dl, retested at 1.01 mg/dl.Sid (b)(6): 0.64 mg/dl, retested at 1.43 mg/dl.Sid (b)(6): 0.39 mg/dl, retested at 1.23 mg/dl.Sid (b)(6): 0.44 mg/dl, retested at 1.23 mg/dl.No impact to patient management was reported.
 
Additional Manufacturer Narrative
All available patient information was included.Additional patient details are not available.Trending review determined no adverse trend for the issue for the product.Historical complaint review determined there is normal complaint activity for the lot number.Device history record review did not show any potential non-conformances, or deviations.Labeling was reviewed and found to adequately address the issue under review.The customer discovered that samples were run while the quality control failed.Rerunning qc then samples has resolved the issue.This indicates the assay is performing as expected.User error may have contributed to the issue.Per the complaint text, the customer received a flag for 21 out of 25 results less than linearity but reported results out.Based on the investigation, no deficiency for lot number 60446uq08 was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTAL BILIRUBIN
Common Device Name
DIAZO COLORIMETRY, BILIRUBIN
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key15094003
Report Number3002809144-2022-00255
Device Sequence Number5323510
Product Code CIG
UDI-Device Identifier00380740174965
UDI-Public00380740174965
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K121985
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Model Number6L45-22
Device Catalogue Number06L45-22
Device Lot Number60446UQ08
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/29/2022
Initial Report FDA Received Date07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/29/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC C4000 INTGR, 02P24-40, C403072
Patient SexUnknown
-
-