(b)(4).Batch #: 341a52.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the cs40b device was received with no apparent damage and with a reload loaded in the device.The reload was a received with all staples protruding, the washer uncut and with the knife recessed below the reload deck.The drivers were noted to be partially advance.The reload was not properly loaded in the device, as the retaining pin was not connected to the coupler and pushrod.These facts indicate that the reload was removed and reinserted incorrectly and/or incompletely after the retainer was removed.The device was tested for functionality with a test reload and it fired, forming all staples and cutting as intended.The cut line and the staple line were completed and the staples meet the staple form release criteria, also, the device opened and closed without any difficulties.The condition of the reload is consistent with the device being partially activated.As a result of the partial firing the lockout was engaged when the device was reopened.Do not fire the instrument unless the closure trigger is properly latched against the handle.The firing trigger must be pulled back completely against the closure trigger to properly fire the instrument.In addition, if the firing sequence is not complete, you could deploy the staples without cutting the washer and forming the staples.It should be noted that if the reload is removed from the device, whether the reload is spent or not, and if the staple retainer has already been removed from the reload, the reload cannot be reloaded into the device.Please reference the instructions for use for the complete guide loading and reloading the instrument.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|