Event summary as reported, when the doctor took a lapsac tissue entrapment pouch, he noticed the outer package was contaminated.The procedure was completed by using a new product.Investigation - evaluation reviews of the complaint history, device history record, instructions for use, and quality control procedures and a visual inspection of the device were conducted during the investigation.The device was returned in a zipped plastic bag without the original packaging for investigation.The product appears unused.One loose piece of red foreign matter was found inside the pouch.A document-based investigation evaluation was performed.No related non-conformances were recorded.No additional complaints were received for this product lot.The device history record review provides objective evidence that the device was manufactured to specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was functionally inspected by quality control and no notable gaps in production or processing controls were noted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.The device is provided with instructions for use which caution, ¿supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not use the product if there is doubt as to whether the product is sterile.¿ based on the available information, cook has concluded that the cause of the complaint is likely manufacturing related.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per a review of risk documentation, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious.Injury if a malfunction occurred.
|